Experience the best in solvent protection with Caron's ductless fume cabinets, where safety meets visibility.
Up to three layers of protection through either general-purpose or solvent-specific organic carbon media
Clearview acrylic construction for visibility and comfort
Significantly lower long-term cost of ownership compared to traditional ducted systems.
| MODEL | BC 0402 | BC 0603 | BC 0804 | BC 1004 | BC 1206 | BC 1506 | PC 0804 | PC 1004 |
| PRE-FILTER | Oiled pre-filter media eliminates particles >5µm to ISO 16890 | |||||||
| MAIN- FILTER (2) | See Filter Selection | |||||||
| MAIN FILTER (3) | Optional | |||||||
| OPENING TYPE |
Flap | Teardrop or Trapezium | Trapezium | Teardrop or Trapezium | ||||
| VOLUME OF AIR |
92m³/hr | 130m³/hr | ||||||
| FACE VELOCITY |
>0.5m/s | |||||||
| FAN TYPE | EBM R3G190-RG15-44 50/60Hz dynamically balanced centrifugal (115V) |
EBM R3G190-RG15 44 50/60Hz dynamically balanced centrifugal, x2 (115V) |
EBM R3G190-RG15-44 50/60Hz dynamically balanced centrifugal (115V) |
|||||
| EBM R3G190-RC05-03 50Hz-dynamically balanced centrifugal (230V) |
EBM R3G190-RC05 03 50 Hz dynamically balanced centrifugal, x2 (230V) |
EBM R3G190-RC05-03 50Hz-dynamically balanced centrifugal (230V) |
||||||
| FAN CONTROL | Automatic, Programmable Nextion TFT Fan & Alarm Control System | |||||||
| AIRFLOW ALARM |
Nextion TFT Fan & Alarm Control System | |||||||
| LIGHTING | Integrated LED Lighting (optional) | Integrated LED Lighting |
Integrated LED Lighting (optional) |
|||||
| SOUND LEVEL | < 50dBA at 1 meter | |||||||
| FINISH EXT | Light Grey, Gloss Finish, Plastic Coating | |||||||
| COMPLIANCE STANDARDS |
BS7989:2001 Specification For Recirculatory Filtration Fume Cupboards |
|||||||
| AFNOR NFX 15-211 Class 1 For Filter Retention Capacity | ||||||||
| AFNOR NFX 15-203 For Containment | ||||||||
| Bs EN 61010:2010 Safety Requirements for Electrical Equipment For Measurement, Control And Laboratory Use |
||||||||
| BS EN 61326:2013 Electrical Equipment for Measurement, Control And Laboratory Use. EMC Requirements. |
||||||||
| Carbon Filter: BS7989:2001, HEPA Filter: EN1822:2009 | ||||||||
Model: Ductless Fume 600mm
Description: Ductless Fume hoods, Single & Dual layer Chemcap Carbon